ETX-1825,multi calibrator,portable multifunction calibrator,Multifunction Calibrator,0.05% Accuracy

ETX-1825,multi calibrator,portable multifunction calibrator,Multifunction Calibrator,0.05% Accuracy
Product Detailed
1.Calibrates a transmitter using the split-screen 2.Manual stepping and automatic stepping and ramping

Features and Benefits

A split-screen display. The upper display allows you to measure volts, current (with loop power), resistance and continuity test. The lower display allows you to measure and source volts, current, resistance temperature detectors (RTD), thermocouples (TC), frequency, and ohms.Calibrates a transmitter using the split-screen.The measure of thermocouple provides automatic reference junction.Automatic and manual cold end temperature compensation for TC.Manual stepping and automatic stepping and ramping.Super-strong protection:The waterproof grade is IP67. Automatic protection against the signal terminals connect to the 220V.

Summary of Source and Measure Functions

Function

Measure

Source

DC V

0~50V (the upper display ± 30V)

0~10V

DC mA

0~24mA (the upper display ± 24mA)

0~24mA

Frequency

1.000Hz ~ 100.00kHz

0.00Hz~20.000kHz

Resistance

0~3200Ω

0~3200Ω

DC mV

0~100mV (the upper display ± 200mV)

0~100mV

RTD

Pt100, Pt1000, Cu50, Cu100

Pt100, Pt1000, Cu50, Cu100

TC

E, J, K, T, B, R, S, N

E, J, K, T, B, R, S, N

Others

Loop supply, Step, Ramp, Dual display

  Specifications

Specifications are based on a one year calibration cycle and apply from +18 ºC to +28ºC unless stated otherwise. All specifications assume a 10 minute warmup period.

DC Voltage Measurement

Range

Maximum measurement range

Resolution

Accuracy ( % of reading + counts)

ETX-1825

ETX-2025

28V

(upper display)

-33V~33V

0.001V

0.05+2

0.025+2

200mV

(upper display)

-200mV~200mV

0.01mV

0.05+2

0.025+2

50V

(lower display)

-1V~60V

0.001V

0.05+2

0.02+2

100mV

(lower display)

-15mV~125mV

0.01mV

0.05+2

0.02+2

-10°C~18°C,+28°C~55°C temperature coefficient,±0.005%FS/°C.

Input resistance: >1MΩ.

  DC mA measurement

Range

Maximum measurement range

Resolution

Accuracy ( % of reading + counts)

ETX-1825

ETX-2025

20mA

(upper display)

-24mA~24mA

0.001mA

0.05+2

0.025+2

20mA

(loop of upper display)

0~24mA

0.001mA

0.05+2

0.025+2

20mA

(lower display)

0~24mA

0.001mA

0.05+2

0.02+2

-10°C~18°C,+28°C~55°C temperature coefficient,±0.005%FS/°C..

Input resistance:<100Ω.

  Resistance Measurement (upper display)

Range Maximum output range Resolution Accuracy ( % of reading + counts)
ETX-1825ETX-2025
400Ω0~440Ω0.1Ω0.05+2
3200Ω420Ω~3300Ω
On-off test 0~200Ω0.05+1
-10°C~18°C,+28°C~55°C temperature coefficient,±0.005%FS/°C.
Maximum load voltage: 20V, equivalent to voltage of 20mA on 1000Ω load resistance.

 Resistance measurement (lower display)

Range Maximum measurement rangeResolution Accuracy(Ω)
ETX-1825ETX-2025
Two-wire, three-wireFour-wireTwo-wire, three-wireFour-wire
400Ω0~440Ω0.01Ω0.250.150.150.1
3200Ω420Ω~3600Ω0.1Ω1.5110.5
-10°C~18°C,+28°C~55°C temperature coefficient,±0.005%FS/°C.
Exciting current during measurement: 400Ω:0.4mA±10%;3200Ω:0.2mA±10%.
Two-wire: Does not include lead resistance.
Three-wire: Assumes matched leads with a total resistance not exceeding 25Ω.